| OTR
Testing Of PET Bottles |
Product
Overview |
Illinois
Instruments’ 8000 Series of analyzers provides an analyzer to suit any
oxygen permeation requirement.
These
economical and precise instruments are simple to use and very versatile.
Some of the features are:
|
|
Widest
measuring range in marketplace (0.008 – 432000 cc/m2/day) |
|
|
Single
sensor to cover the entire range |
|
|
Precise
relative humidity control of 8001 |
|
|
Precise
temperature control |
|
|
No
special gas mixtures required |
|
|
Carrier
gas and test gas controlled by mass flow controller |
|
|
Calibrate
to certified gas or a reference film |
|
|
Test
both packages and films on same instrument (except 8200 and 8501) |
|
|
Fast
changeover from wet to dry samples |
Principle
Of Operation:
The
8000 series of oxygen permeation analyzers utilizes our proprietary Coulometric
sensor technology to detect oxygen transmission rates (OTR) through flat
film barriers, PET bottles and various kinds of packages
Samples
are clamped or attached to a diffusion chamber. Pure oxygen (99.9%(
is then introduced into the upper half of the chamber while an oxygen-free
carrier gas flows through the lower half.
Molecules
of oxygen diffusing through the sample into the lower chamber are conveyed
to the sensor by the carrier gas. This allows a direct measurement
of the oxygen without using complex extrapolations. Oxygen transmission
rate of the test sample is display as either cc/m2/day, cc/100in2/day or
cc/package/day
Model
8001 is
designed specifically for trace oxygen permeation analysis of packaging
barrier films. The 8001 is both modular and extremely flexible. The
Model 8001 provides the widest oxygen permeation measurement range on the
market today. It has standard features such as precise temperature
and precise humidity control and both dry and wet samples may be run. Both
films and packages may be tested. It is modular in design and up
to 5 expansion modules may be added.
.
Model
8002 Oxygen
Permeation Analyzer has all the features of the 8001 with the exception
of precise humidity control. The 8002 is designed for those laboratories
that only require the running of dry and wet samples. The full range
of temperature is available and samples are run either dry or saturated
(assumed 100%).
Model
8003 Oxygen Permeation
Analyzer is for the laboratory that does not have a requirement to test
wet samples. This instrument is for dry analysis only but has all
the other features of the Model 8001.
Model
8200 Oxygen Permeation
Analyzer is designed for those companies that test only packages, such
as PET bottles. It has the same accuracy and ease of operation as
the 8001, but it is an economical choice for PET bottle analysis.
PET bottles are attached to the instrument, and an oxygen-free carrier
gas flows through the inside of the bottle picking up the oxygen that permeates
through the PET material, enabling simple and precise measurement.
Model
8501 offers a cost
effective solution for those who need accurate, uncomplicated measurement
of oxygen transmission rates (OTR) through permeable films
|